False-negative results from point-of-care qualitative human chorionic gonadotropin (hCG) devices caused by excess hCGbeta core fragment vary with device lot number.
Gronowski, Ann M
False-negative results from point-of-care qualitative human chorionic gonadotropin (hCG) devices caused by excess hCGbeta core fragment vary with device lot number. [electronic resource] - Clinical chemistry Oct 2009 - 1885-6 p. digital
Publication Type: Case Reports; Letter
1530-8561
10.1373/clinchem.2009.133280 doi
Adolescent
Chorionic Gonadotropin, beta Subunit, Human--standards
Equipment Failure
False Negative Reactions
Female
Humans
Peptide Fragments--standards
Point-of-Care Systems
Pregnancy
Reagent Kits, Diagnostic
Reference Standards
Young Adult